The relationship between aortic flow and pressure is described by a three-element model of the arterial input impedance, including continuous correction for variations in the diameter and the compliance of the aorta (Modelflow). We computed the aortic flow from arterial pressure by this model, and evaluated whether, under orthostatic stress, flow may be derived from both an invasive and a non-invasive determination of arterial pressure. In 10 young adults, Modelflow stroke volume (MFSV) was computed from both intra-brachial arterial pressure (IAP) and non-invasive finger pressure (FINAP) measurements. For comparison, a computer-controlled series of four thermodilution estimates (thermodilution-determined stroke volume; TDSV) were averaged for the following positions: supine, standing, head-down tilt at 20 ° (HDT20) and head-up tilt at 30 ° and 70 ° (HUT30 and HUT70 respectively). Data from one subject were discarded due to malfunctioning thermodilution injections. A total of 155 recordings from 160 series were available for comparison. The supine TDSV of 113±13 ml (mean±S.D.) dropped by 40% to 68±14 ml during standing, by 24% to 86±12 ml during HUT30, and by 51% to 55±15 ml during HUT70. During HDT20, TDSV was 114±13 ml. MFSV for IAP underestimated TDSV during HDT20 (-6±6 ml; P < 0.05), but that for FINAP did not (-4±7 ml; not significant). For HUT70 and standing, MFSV for IAP overestimated TDSV by 11±10 ml (HUT70; P < 0.01) and 12±9 ml (standing; P < 0.01). However, the offset of MFSV for FINAP was not significant for either HUT70 (3±8 ml) or standing (3±9 ml). In conclusion, due to orthostasis, changes in the aortic transmural pressure may lead to an offset in MFSV from IAP. However, Modelflow correctly calculated aortic flow from non-invasively determined finger pressure during orthostasis.
Skip Nav Destination
Article navigation
Research Article|
July 21 1999
Continuous stroke volume monitoring by modelling flow from non-invasive measurement of arterial pressure in humans under orthostatic stress
Mark P. M. HARMS;
Mark P. M. HARMS
*Department of Internal Medicine, Academic Medical Center, Cardiovascular Research Institute, Amsterdam, The Netherlands
Search for other works by this author on:
Karel H. WESSELING;
Karel H. WESSELING
†Netherlands Organization for Applied Scientific Research, TNO Biomedical Instrumentation, Amsterdam, The Netherlands
Search for other works by this author on:
Frank POTT;
Frank POTT
‡Department of Anesthesia, Rigshospitalet, The Copenhagen Muscle Research Center, Copenhagen, Denmark
Search for other works by this author on:
Morten JENSTRUP;
Morten JENSTRUP
‡Department of Anesthesia, Rigshospitalet, The Copenhagen Muscle Research Center, Copenhagen, Denmark
Search for other works by this author on:
Jeroen VAN GOUDOEVER;
Jeroen VAN GOUDOEVER
†Netherlands Organization for Applied Scientific Research, TNO Biomedical Instrumentation, Amsterdam, The Netherlands
Search for other works by this author on:
Niels H. SECHER;
Niels H. SECHER
‡Department of Anesthesia, Rigshospitalet, The Copenhagen Muscle Research Center, Copenhagen, Denmark
Search for other works by this author on:
Johannes J. VAN LIESHOUT
†Netherlands Organization for Applied Scientific Research, TNO Biomedical Instrumentation, Amsterdam, The Netherlands
Correspondence: Dr J. J. van Lieshout, Department of Internal Medicine, Room F4–264, Academic Medical Center, University of Amsterdam, P.O. Box 22700, 1100 DE Amsterdam, The Netherlands.
Search for other works by this author on:
Publisher: Portland Press Ltd
Received:
February 12 1999
Revision Received:
April 15 1999
Accepted:
May 23 1999
Online ISSN: 1470-8736
Print ISSN: 0143-5221
The Biochemical Society and the Medical Research Society © 1999
1999
Clin Sci (Lond) (1999) 97 (3): 291–301.
Article history
Received:
February 12 1999
Revision Received:
April 15 1999
Accepted:
May 23 1999
Citation
Mark P. M. HARMS, Karel H. WESSELING, Frank POTT, Morten JENSTRUP, Jeroen VAN GOUDOEVER, Niels H. SECHER, Johannes J. VAN LIESHOUT; Continuous stroke volume monitoring by modelling flow from non-invasive measurement of arterial pressure in humans under orthostatic stress. Clin Sci (Lond) 1 September 1999; 97 (3): 291–301. doi: https://doi.org/10.1042/cs0970291
Download citation file:
Sign in
Don't already have an account? Register
Sign in to your personal account
You could not be signed in. Please check your email address / username and password and try again.
Could not validate captcha. Please try again.